Memmert ICH260 Stability Climate Chamber | 9.04 cu. ft.
Item#: ICH260115VICH 260 Memmert stability climate chambers guarantee the perfect atmosphere for climate and temperature tests. Unparalleled homogeneity of temperature and humidity.
$21,468.00
Supplier: Memmert
Tags: Stability Chambers, Temperature Test Chambers- Volume: 9.04 cu. ft. | 256 L
- Interior Dimensions W x H x D in (mm): 25.20" x 31.50" x 19.67" (640 x 800 x 500 mm)
- Exterior Dimensions W x H x D in (mm): 32.44" x 61.10" x 28.90" (824 x 1552 x 734 mm)
- Housing: rear zinc-plated steel
- Installation: on lockable castors
- Max. number of internals: 9
- Max. loading of chamber: 440.9 lbs (200 kgs)
- Max. loading per internal: 44.09 lbs (20 kgs)
- Electrical data: 115 V, 50/60 Hz approx. 1450 W (230V model available, please inquire)
Temperature Specs:
- Setting temperature range: -10 to +60 °C
- Working temperature range without humidity: -10°C up to 60°C
- Working temperature range with humidity: 10°C up to 60°C
- Setting accuracy temperature: 0.1°C
- Temperature sensor: 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
Humidity Specs:
- Humidity supply with distilled water from external tank by self-priming pump
- Humidification by hot steam generator
- Active humidifying and de-humidifying adjustable from 10-80 % rh with digital display of relative humidity - resolution of display 0.1 %, setting accuracy 0.5 %
- Stainless Steel Exterior: Scratch resistant, hygienic and durable.
- Warranty: A limited 2 year warranty is included with the option to purchase an extra year.
- ControlCOCKPIT: TwinDISPLAY. Adaptive multi-functional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
- Language Setting: German, English, Spanish, French, Polish, Czech, Hungarian
- Adjustable Parameters: Temperature (Celsius or Fahrenheit), relative humidity, program time, time zones, summertime/wintertime
- Timer: Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
- Function SetpointWAIT: The process time does not start until the set temperature is reached
- Calibration: Three freely selectable values each, temperature/humidity/CO2
- Temperature control: mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 10°C above nominal temperature
- AutoSAFETY: additionally integrated over- and under temperature protection “ASF”, automatically following the setpoint value at a preset tolerance range, alarm in case of over- or under temperature, heating is switched off in case of overtemperature, the compressor in case of under temperature
- Alarm: with visual and acoustic alarm in case of over-/under temperature and under humidity, open door and empty water tank
- Temperature control: over- and under temperature monitor TWW, protection class 3.3 or adjustable temperature limiter TWB, protection class 2, adjustable at the controller
- Autodiagnostic system: integral fault diagnostics for temperature and humidity control
Request a Proposal
Related Products
- Item #: HPP110ECO $18,055.00
- Item #: ICH110L115V $18,040.00
- Item #: ICH110115V $14,288.00
- Item #: HPP260ECO $23,773.00
- Item #: ICH260L115V $27,158.00
- Item #: HPP410ECO $28,955.00
-
Item #: NSRI231WSG-0H
$27,217.00$23,134.00 -
Item #: NSRI231WSW-0H
$26,502.00$22,527.00