800-521-0754

Request Information

Request Info

CLOSE

FARRAR™ Ultra-Low Chamber Pallet Storage Freezer | -20°C to -80°C

Item#: ULC-P

FARRAR™ Ultra-Low Chamber (ULC) and 'Pallet' Ultra-Low Chamber (ULC-P) are the only -20°C to -80°C, forced-air, ultra-low-temperature chambers on the market. Purpose-built for life sciences, our ultra-low temp bulk storage solutions are unmatched in temperature uniformity and recovery.

Request a Quote
  • Total Volume: 311 cu.ft / 8,438 l
  • Door Opening Size (W x H)/QTY: 41” x 87” / 2 (1041.4 x 2209.8 mm)
  • Pallet Capacity (Euro Pallet):
    • 2 - 32” x 47” x 50” (W x D x H) (2 -813 x 1194 x 1270 mm)
    • 4 - 32” x 47” x 36” (W x D x H) (4 - 813x 1194 x 914.4 mm)
  • Area Footprint: 67.5 sq.ft./ 6.28 m2
  • Interior Dimensions (W x D x H): 110” x 54” x 90” (2794 x 1371 x 2286 mm)
  • Exterior Dimensions (W x D x H): 122” x 66” x 143” (3098 x 1676 x 3632 mm)
  • Unit Weight: 3,900 lbs. (1,769 Kg)
  • Shipping Weight: 4,500 lbs. (2,132 Kg)
  • Application: Pharmaceutical / CDMO’s and CRO’s
  • Chamber Volume: 311 Cu. Ft. (8,806 L)
  • Temperature Range: -20°C to –80°C
  • Electrical: 208/230 VAC, 1 Ø, 60Hz, 60Hz
  • Circuit Breaker: 2 x 30 amp
  • Ambient Operating Temperature: +18°C to +30°C
  • Uniformity Air Temperature: +/- 3.0°C (empty chamber)
  • Uniformity Product Temperature: +/- 1.0°C
  • Evaporator/QTY: Forced - Air / 2
  • Condenser/QTY: Forced - Air Cooled / 2
  • Metering Device: Electronic
  • High Stage/ Low Stage: R-449A / R-508B
  • Air Cooled Condenser Requirements: +18°C to +30°C
  • Water Cooled Condenser Requirements:
    • Tower Water 85°F (29.5°C), Max. flow rate 7gpm (26.5 l/min)
    • Chilled Water 45°F (7.2°C), Max. flow rate 3.5gpm (13.25 l/min)
  • Temperature Controller/QTY: Microprocessor/ 2
  • Display/QTY: LCD/ 2
  • Sensor/QTY: RTD PT100 Din A/ 2
  • Alarms: High/Low temperature, door ajar, power failure, remote alarm contacts
  • Dry Contacts: Standard (C, NO, NC)
  • Warranty: 1 year parts & labor
  • Specifically designed for Pharmaceutical, CDMO’s and CRO’s
  • Largest volume space available for a -80°C freezer
  • Temperature operating range -20°C to -80°C
  • Smallest foot print with largest storage volume
  • Forced-air circulation provides fastest temperature recovery from door opening
  • Fully redundant systems offer peace of mind (display, electrical, refrigeration and controls)
  • Optional racking/shelving system available

Options:

  • Air/water cooling
  • IQOQ performance
  • Certified setup/startup
  • Hydraulic Lift movement set
  • 2nd year warranty
  • Shelves or SS racking (ULC-P only)

Request a Proposal