Memmert UF110plus Universal Forced Circulation Lab Oven, 115V | 3.81 Cu. ft.
Item#: UF110Plus115VThe Memmert UF110plus 3.81 cu. ft. forced circulation laboratory oven provides precise, uniform heating for laboratory applications up to 300°C. Adjustable forced-air circulation promotes consistent chamber conditions, while TwinDISPLAY ControlCOCKPIT technology provides precise PID temperature control. Dual Pt100 sensors continuously monitor temperature and provide redundancy if a sensor error occurs.
HeatBALANCE and SetpointWAIT functions support controlled heating and process timing, while AtmoCONTROL software enables program management through Ethernet or USB. AutoSAFETY, overtemperature protection, visual and acoustic alarms, and automatic diagnostics support safe operation. The 115V oven also includes preheated fresh-air intake, adjustable fan speed, and an electronically controlled air flap.
$5,668.00
- Working temperature range: at least 5 (UN/UNplus/UNm/UNmplus) or 10 (UF/UFplus/UFm/UFmplus) above ambient temperature to +300 °C
- Setting accuracy temperature: up to 99.9 °C: 0.1 / from 100 °C: 0.5
- Setting temperature range: +20 to +300 °C
- Temperature sensor: 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
- Fan: forced air circulation by quiet air turbine, adjustable in 10 % steps for each segment individually
- Fresh air: Admixture of pre-heated fresh air by electronically adjustable air flap
- Vent: vent connection with restrictor flap
- Works calibration certificate: Calibration at +160°C
- Door: fully insulated stainless steel doors with 2-point locking (compression door lock)
- Internals: 2 stainless steel grid(s), electropolished
- Interior Dimensions (W x H x D): 22.05" x 18.90" x 15.75"
- Interior: easy-to-clean interior, made of stainless steel, reinforced by deep drawn ribbing with integrated and protected large-area heating on four sides
- Volume: 3.81 Cu.ft.
- Max. number of internals: 5
- Max. loading of chamber: 385.81 lbs.
- Max. loading per internal: 44 lbs.
- Textured stainless steel casing
- Dimensions (W x H x D): 29.33" x 34.01" x 22.99" (d +2.20" door handle)
- Housing: rear zinc-plated steel
- Voltage Electrical load: 115 V, 50/60 Hz approx. 1800 W (230V available)
- Set Up: The distance between the wall and the rear of the appliance must be at least 6". The clearance from the ceiling must not be less than 8" and the side clearance from walls or nearby appliances must not be less than 2".
- Altitude of installation: max. 2,000 m above sea level
- Ambient temperature: +5 °C to +40 °C
- Humidity rh: max. 80 %, non-condensing
- Overvoltage category: II
- Pollution degree: 2
- Transport information: The appliances must be transported upright
- Shipping Dimensions (including carton) (w x h x d): 32.68" x 41.34" x 31.50"
- Net weight: approx. 163.14 lbs.
- Shipping Gross weight (including carton): approx. 218.26 lbs.
- ControlCOCKPIT: TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-color displays.
- Timer: Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
- Function HeatBALANCE: adapting the distribution of the heating performance of the upper and lower heating circuit from -50 % to +50 %
- Function SetpointWAIT: the process time does not start until the set temperature is reached
- Calibration: three freely selectable temperature values
- Adjustable Parameters: temperature (Celsius or Fahrenheit), fan speed, air flap position, program time, time zones, summertime/wintertime
- Documentation: program stored in case of power failure
- Programming: AtmoCONTROL software on a USB stick for programming, managing and transferring programs via Ethernet interface or USB port
- Temperature control: mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature
- Temperature control: overtemperature monitor TWW, protection class 3.1 or adjustable temperature limiter TWB, protection class 2, selectable on display
- AutoSAFETY: additionally integrated over- and under temperature monitor "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or under temperature, heating is switched off in case of overtemperature
- Auto diagnostic system: for fault analysis
- Alarm: visual and acoustic
Request a Proposal
Related Products
- Item #: UF30115V $2,543.00
- Item #: UF30Plus115V $3,798.00
- Item #: UF55Plus115V $4,457.00
- Item #: UF55115V $3,064.00
- Item #: UF75Plus115V $4,914.00
- Item #: UF75115V $3,621.00
- Item #: UF110115V $4,232.00
- Item #: UF160Plus115V $7,064.00