800-521-0754

Request Information

Request Info

CLOSE

Memmert UN55 Universal Forced Circulation Lab Oven, 115V | 1.9 Cu. ft.

Item#: UF55115V

Memmert UN55 universal laboratory oven is a 1.9 cu. ft. natural convection oven designed for precision heating, drying, conditioning, and material testing applications in research, pharmaceutical, and industrial laboratories. The unit provides temperature control from ambient +10°C to 300°C using a PID microprocessor controller and natural convection airflow for gentle, uniform heating of temperature-sensitive materials. A high-definition ControlCOCKPIT display supports advanced programming, calibration, and monitoring, while the stainless-steel interior offers durability and easy cleaning. Integrated overtemperature protection and auto-diagnostic capabilities help ensure reliable and repeatable thermal processing performance.

$3,064.00

Supplier:
  • Volume Approx: 1.9 ft/ 53 L
  • Interior Width: 15.7.0 in / 400 mm
  • Interior Height: 15.7.0 in / 400 mm
  • Interior Depth: 13.0 in 330 mm
  • Number of Stainless Steel Grids (Standard Equipment): 1
  • Maximum Number of Grids: 4
  • Maximum Loading per Grid: 44 lb / 20 kg
  • Maximum Loading of Chamber: 176 lb / 80 kg
  • Net weight approx.: 126.67 lbs
  • Gross weight carton approx.: 167.55 lbs
  • 115 V (+ / - 10%), 50/60 Hz / 14.8 amps
  • 3 Year Warranty
  • Temperature: 1 Pt100 sensor DIN class A in 4-wire-circuit
  • Set temperature range: min.10°C above ambient up to +300°C
  • Resolution of display for setpoint values 0.1°C up to 99.9°C, 0.5°C from 100°C and for actual values 0.1°C (LED)
  • ControlCOCKPIT: adaptive multifunctional digital PID-microprocessor controller with high-definition TFT-color display
  • Timer: Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
  • Adjustable Parameters: temperature (Celsius or Fahrenheit), air flap position, program time, time zones, daylight savings time
  • Function SetpointWAIT: the process time does not start until the set temperature is reached
  • Language Setting: German, English, Spanish, and French
  • Calibration: three freely selectable temperature values
  • Temperature control: adjustable electronic overtemperature monitor and mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature
  • Auto diagnostic system: for fault analysis

Request a Proposal