800-521-0754

Request Information

CLOSE

NuAire In-VitroCell ES NU-8645 5.65 cu. ft. (160L) Water Jacketed CO2 Incubator

Item#: NU-8645

The In-VitroCell NU-8645 CO2 Incubator uses a water jacket to control and maintain the temperature in the growth chamber. In addition, model NU-8645 features an oxygen suppression system to control O2 levels between 0.5 to 21 percent for mammalian tissue culture research. The NU-8645 uses a water pan and convection to generate up to 95% humidity levels. Monitor relative humidity in the chamber through an RH sensor and display the percentage on the incubator touchscreen controls. In-VitroCell NU-8645 provides uniform and stable growth conditions for cell culture research.

Request a Quote
  • Shipping Inches (mm): 31 (787) W x 49 (1245) H x 33 (838) D
  • Access Port Location: Right Side
  • Air Cleanliness (ISO 14644): ISO-Class-5
  • Chamber Air Circulation Method: Air Pump
  • Chamber Air Filter: 99.97% Efficiency @ ≥ 0.3 Micron Capsule Filter Located in Sensor Bay
  • Chamber Heater(s) Location: Bottom of Chamber
  • Chamber Jacket Technology: Water Jacket, 18 Gallon (68-Liter)
  • CO2 Accuracy: ± 0.1%
  • CO2 Control Logic: Fixed Algorithm/Manual, Environmental Adaptable
  • CO2 Display Resolution: 0.1%
  • CO2 Range: 0.1 to 20%, 5% Default Set Point
  • CO2 Recovery: Up to 5% ± 0.2% / -0.5 in 5 minutes Average
  • CO2 Sensor Type: Infrared (NDIR)
  • Control System: NuTouch
  • Data Output Features: RS-485, USB, 4-20mA
  • Door and Perimeter Heater Control Logic: Base Duty Cycle set to chamber temperature Set-Point, Proportional 0-100% [manually adjustable
  • Electrical Configuration: 115V50Hz, 230V5060Hz
  • Exterior Depth: 27” (685mm)
  • Exterior Door Seal: Magnetic Outer Door Gasket
  • Exterior Height Inches (mm): 37.75 (958)
  • Exterior Material: 16 Gauge Steel with Textured Polyurethane Finish
  • Exterior Width Inches (mm): 25.625 (649)
  • Footprint Depth inches (mm): 17.5 (445)
  • Footprint Width inches (mm): 22 (557)
  • In-line Gas Filters: 99.97% Efficiency @ ≥ 0.3 Micron Disk Filters for Gas & Air Supplies, Disk Filter for CO2 Sensor
  • Segmented Doors: Optional Set of 4 Inner Doors
  • Inner Doors: One Standard, Sectional optional
  • Insulation: Insulation with R5.0 Rating
  • Interior Depth Inches (mm): 20.625 (524)
  • Interior Door Seal: Silicone Rubber Gasket
  • Interior Height Inches (mm): 24 (611)
  • Interior Material: Polished Type 304L 16 Gauge Stainless Steel
  • Interior Volume Cubic Feet (Liters): 5.65 (160)
  • Interior Width Inches (mm): 20.375 (518)
  • Net Weight lbs (kg): 403 (183) includes full water jacket and shelves
  • O2 Accuracy: ± 0.25%
  • O2 Control Logic: Fixed algorithm/ manual environmental adaptable
  • O2 Display Resolution: 0.1%
  • O2 Measuring Range: 0 - 25%
  • O2 Set Point Range: 0.5 -21.0% (ambient) (21.0% Default)
  • O2 Recovery: 5.0 ± 0.5% twenty minutes on Average
  • O2 Sensor Type: Zirconia Ceramic
  • Power Cord Length Foot (Meter): 8 (2.5)
  • Remote Alarm Terminals: Standard
  • RH Accuracy: Not available
  • RH Water Pan Capacity: 2L (Recommended Fill 1.5L)
  • Sample Port Location: Inner Glass Door
  • Shelf Capacity lbs (kg): 25 lbs per Shelf / 125 lbs per Incubator
  • Shelf Depth Inches (mm): 18 3/4 (476)
  • Shelf Material: Stainless Steel or Copper Alloy
  • Shelf Width Inches (mm): 18 (457)
  • Shelves: 4 Shelves Supplied, 16 Shelves Maximum Capacity
  • Shipping Weight lbs (kg): 370 (168)
  • Decontamination Cycle: Not Available
  • Temperature Accuracy °C: ± 0.1%
  • Temperature Control: Water Jacket
  • Temperature Control Range: 5°C above ambient (to a 30°C max ambient) to 55°C
  • Temperature Display Resolution : 0.1°C
  • Temperature Recovery: 12°C/min. on Average
  • Temperature Sensor Type: Precision Integrated Circuit
  • Temperature Set-Point Range: 5°C to 55°C (37.0°C Default)
  • Temperature Uniformity °C: ± 0.20°C @ 37°C
  • Water Pan Depth Inches (mm): 1.5 (38)
  • Water Pan Length Inches (mm): 12 (305)
  • Water Pan Width Inches (mm): 10 (254)
  • NuTouch Electronic Control System
  • 100% Stainless Steel Coved Interior Chamber
  • Dual Temperature Sensor Probes
  • Infrared (IR) CO2 Sensor
  • O2 Control System (NU-8631)
  • Four (4) Stainless Steel Shelves
  • Eight (8) Stainless Steel Shelf Guides
  • Four (4) Wall Brackets
  • Right Hinged Door Swing
  • Remote Alarm Output Contacts
  • 4 to 20 mA Analog Output
  • RS-485 Communication
  • USB Port
  • CO2 Sample Port
  • Adjustable Leg Levelers
  • Access Port and Plug with Breather Holes
  • One (1) Water Pan
  • One (1) Electrical Cord
  • Warranty: 3 Years Parts & Labor (5 Years IR Sensor Parts Only)

Electrical Requirements

  • Startup Power: 475 watts
  • Running Power: 250 watts
  • Heat Rejected: 14 BTU / min.

Utility Connections

  • Gas Connections: 0.25 in.
    (6.3 mm) Tubing Connections
  • Gas Input Pressure: 20 PSIG
    (1.4 BAR) Input Pressures
    Maximum. Two-Stage Gas
    Regulators Required.

In-VitroCell CO2 Incubator Product Brochure

Download

NU-8600 & NU-8631 CO2 Incubator Manual

Download

Request a Proposal