LabRepCo CliniCool© Series 4.2 Cu. Ft. Undercounter NSF Certified Pharmacy/Vaccine Freezer | Built-In | Stainless Steel
Item#: LHP-5-UFMBSS-PH-NSFThe LabRepCo CliniCool© Series LHP-5-UFMBSS-PH-NSF is a 4.2 cu. ft. built-in medical freezer certified to NSF/ANSI 456 for vaccine storage. It operates from -15°C to -28°C with microprocessor temperature control. The stainless-steel exterior, solid locking door, and two fixed shelves provide secure undercounter storage. Audible and visual temperature alarms and remote contacts support critical temperature monitoring. A calibrated probe and access port provide additional monitoring capabilities for stored vaccines and pharmaceuticals. The freezer uses R600a refrigerant and is compatible with Pyxis®, Omnicell®, and AcuDose RX® systems.
Please note this product ships freight
What does NSF certified vaccine storage mean? Click here to learn more
$2,660.00 $2,048.20
- Temperature setpoint range: -15°C to -28°C (Controller settings must remain unaltered to ensure thermal performance compliant with NSF/ANSI 456 Standard for Vaccine Storage requirements)
- Operational environment: Indoor use only. Optimal operating range: +18°C to +26°C (+65°F to +78°F), 70% RH
- Exterior Dimensions (WxDxH): 23 3/4" x 24 1/2" x 33 3/8"
- Interior Dimensions (WxDxH): 19" x 17 1/4" x 21"
- Storage capacity: 4.2 cu. ft. gross volume
- Door: One swing solid door, self-closing, right hinged, non-reversible, magnetic sealed gasket, keyed lock
- Shelves: Two shelves, fixed
- Mounting: Low profile roller wheels and leveling legs
- Interior lighting: N/A
- External probe access: Rear wall port (3/4") dia.
- Insulation: Cabinet is foamed-in-place with EPA compliant high density urethane foam
- Exterior materials: Stainless Steel
- Access control: Pyxis®, Omnicell® and AcuDose RX® compatible
- General warranty: Two (2) years parts and labor warranty, excluding display probe calibration
- Compressor warranty: Five (5) years compressor warranty
- Product Weight: 100 lbs.
- Shipping Weight: 132 lbs.
- Rated Amperage: 1.5 Amps
- Power Plug/Power Cord: NEMA 5-15 plug, 8 to 10 ft typical, conforms to UL471 requirements, Vaccine storage power cord warning label
- Facility Electrical Requirement: 110-120V AC: 15 A (minimum)
- Agency Listing and Certification: Certified in accordance with the NSF/ANSI 456 Standard for Vaccine Storage. UL, C-UL, ETL, C-ETL listed (either single or dual agency listings) and certified to UL471 standard, hydrocarbon refrigerant safety.
- Included Accessories:
- Pharmacy refrigerator/freezer toolkit and temperature logs
Performance
- Uniformity (Cabinet air): +/- 2.6°C
- Stability (Cabinet air): +/- 2.1°C
- Maximum temperature variation (Cabinet air): +/- 2.9°C
- Temperature rise after an after 8 sec door openings: Temperature did not exceed -17.5°C at any probe for all required NSF/ANSI 456 testing protocols
- Recovery after 3 min door opening: All probes recover to under -15°C within 8.2 min.
- Energy consumption: 0.97 KWh/day
- Average heat rejection: 1.97 KWh/day (224 BTU/h)
- Noise pressure level (dBA): 49 or less installed
- Pull down time to nominal operating temp: 51 min
Note: The CDC requires a data logger or temperature monitoring device to consistently record live temperature readings from the inside of a refrigerator or freezer when storing all types of vaccines. Constant monitoring of vaccines at recommended temperatures is essential for maintaining the efficacy of a vaccine. Failing to do so can lead to wasted vaccines or ineffective vaccines being administered to patients. Please follow the CDC requirements for Vaccine Storage & the COVID-19 Vaccination Program found here.
Controller, Configuration, Alarms and Monitoring
- Controller technology: Parametric, microprocessor, LED display with 0.1°C resolution
- Temperature setpoint range: -15°C to -28°C (Controller settings must remain unaltered to ensure thermal performance compliant with NSF/ANSI 456 Standard for Vaccine Storage requirements)
- Display probe: Calibrated, stainless steel
- External alarm connection:
- State switching remote alarm contacts.
- Visual and audible indicators.
- Alarms: High / Low temperature, compliant with alarm requirements defined in the NSF/ANSI 456 Standard for Vaccine Storage
- Simulator ballast: glass bead thermal media
Refrigeration System
- Compressor: Hermetic, high performance
- Refrigerant: EPA SNAP compliant, R600a, Isobutane
- Condenser: Hybrid fin and tube with low noise fan
- Evaporator: Integrated shelf evaporator design
- Defrost: Manual
-

NIST Traceable® High-Accuracy Refrigerator/Freezer Thermometer (1 Probe)
Item #: LABC3-4238$142.00$134.90 -

Custom UPS Battery Backup Systems
Item #: BATTERYBACKUP -

NIST Traceable® Memory-Loc™ Refrigerator/Freezer Data Logging Thermometer for Vaccine Storage (1 Probe)
Item #: LABC3-6440$245.25$232.99
Request a Proposal
Related Products
-
Item #: LHP-5-UFMB-PHNSF
$2,342.00$1,803.34 -
Item #: LHP-5-UFMBPHNSF-ADA
$2,342.00$1,803.34 -
Item #: LHP-2-FM-PHNSF*
$2,200.00$1,694.00 -
Item #: PF-L5181W-PA
$3,608.00$2,381.00 -
Item #: MDF-MU339HL-PA
$7,689.00$5,075.00 - Item #: MDF-U731M-PA
-
Item #: LHP-5-UFMB-PH
$2,226.00$1,714.02 -
Item #: FZR5PLP-TR-00-00
$16,210.00$11,347.00